When a LFB recovery fails the method acceptance criteria, must all samples prepared together in the same batch be re-prepared at all times, or only if the source of the failure is traced to a source other than the spiking of the sample?

In any method, if a QC check such as a LFB fails, the first thing to do is re-run that QC check. Occasionally, an instrument will just have a hiccup (for example, an autosampler withdraws only a partial volume). Confirm by re-running the QC check. If it still fails, then prepare a new QC check and evaluate it. That lets you know if you might have prepared the initial QC check incorrectly. If it fails, then you will need to cast a wider net to uncover the problem and that may entail re-preparing the entire batch.