Is it acceptable to use laboratory generated control limits that extend beyond limits referenced in analytical drinking water methods?

No, the initial QC criteria in a method are intended to be the starting point. As a laboratory becomes more proficient in running a method, control limits should tighten, not become wider. The recovery criterion specified in a method is considered the maximum tolerance limit that cannot be exceeded.